TL2-A Instructions for Use 

    

 
 

The TL2-A is an affordable high precision digital thermometer. It has been designed for use in both laboratory 
and industrial applications. Additional video tutorials and information are included on the software USB flash 
drive. Features of the TL2-A are as follows: 
 
Configurable RTD and SPRT combinations (100

 or 200

)

 

 

Available 2 channel temperature measurement 

 

0.001 degree resolution (0.0001 for 

 

Implements ITS-90 or Callendar-Van Dusen calibration coefficients 

 

Displays temperature in Celsius, Fahrenheit, Kelvin or Ohms 

 

Backlit, Large LCD character display 

 

USB Type ‘A’ PC interface 

 

USB Type ‘B’ flash memory interface 

 

User programmable relay output interface (Temperature triggered) 

 

Detachable AC power cord 

 

IP50 enclosure rating 

 

Adjustable Zero Offset correction 

 
Cautions 
The TL2-A and associated temperature probes are sensitive devices and should be handled with care. Care 
should be taken not to submit the temperature probes to mechanical shock, over-heating or exposure to fluids 
at the probe to cable interface. Overheating the probes beyond their calibrated temperature range, damage 
and/or contamination may cause sensor drift and instability.  
 
Authorized Service 
 
Service and Calibration should only be performed by ThermoProbe, Inc. or an authorized distributor. 
Contact ThermoProbe for questions about setting up this software or using the TL2-A. 
 
USER INTERFACE FRONT 
 
 

 

 

 
 

Power Button 

Select Button 

Down Button 

Channel Indicators 

Up Button 

Содержание TL2-A

Страница 1: ...memory interface User programmable relay output interface Temperature triggered Detachable AC power cord IP50 enclosure rating Adjustable Zero Offset correction Cautions The TL2 A and associated tempe...

Страница 2: ...for channel 1 and channel 2 navigates the menu display next previous menu item Also adjusts the zero correction when in the Zero Adjust Mode Channel Indication Marker indicates which channel is curre...

Страница 3: ...elsius F Fahrenheit AbS Kelvin o Ohms Hds Hundredths tHoU Thousandths UPdAtE Communication mode for coefficient update FLASH PC rAtE StArt StoP donE Select to Exit Menu 1 sec 10 sec 30 sec 60 sec 1 hr...

Страница 4: ...is not detected 4 The up down arrow indicator will blink on the LCD display to indicate that logging is in process 5 To stop the data logging select FL StoP from the menu Wait for the FL DonE message...

Страница 5: ...e provided screw terminal header can be wired in two configurations Terminals 1 2 are normally closed NC and will turn OFF open when the relay is set Terminals 1 3 are normally open NO and will turn O...

Страница 6: ...isplay will show the zero correction for the selected channel and then return to temperature display mode 5 To save the zero correction the Power Button must be pressed This will set the unit to Stand...

Страница 7: ...lick Next 8 On the next screen click Install from a list or specific location Advanced and click Next 9 On the next screen select Search for the best driver in these locations Check the Search removab...

Страница 8: ...5 JK 10 The wizard will complete the installation of the software NOTE If your PC should start a new Found New Hardware Wizard it may be necessary to repeat the above steps to complete the driver inst...

Страница 9: ...Coefficient Entry The Sensor Data software included with the TL2 A USB flash drive allows the user to input sensor coefficients Install the software by browsing to the SensorData folder on the drive a...

Страница 10: ...ction screen A window will open a message stating TL2 A data successfully transferred Select OK to clear this window Once completed the Calibration Data screen will appear 6 Enter the unit serial numb...

Страница 11: ...L2 A Note Hyperterminal is no longer included with Windows 7 systems A copy of Hyperterminal is included with the TL2 A software USB flash drive and can be run from that location Click start Accessori...

Страница 12: ...12 06 2015 JK Select the Com Port for the TL2 A and click OK NOTE See Finding Com Port Section if you don t know which com port to select Set the Port Settings as shown and click OK...

Страница 13: ...w a and wait for a command Toggle Prompt off cr Resume sending temperatures Set Date D YY MM DD cr Set the TL2 A s real time clock s date Set Time T HH MM SS cr Set the TL2 A s real time clock s time...

Страница 14: ...Response failure Rate Format R X X 1 10 30 60 3600 Poll cr lf Remarks The brackets contain the options separated by commas e g R 10 cr R poll cr A lower case r or the word Rate or rate can be used En...

Страница 15: ...he text as follows Date YY YY MM DD 3 To change the time enter the text as follows Time HH MM SS Saving a file from HyperTerminal 1 From the HyperTerminal menu bar click Transfer Capture Text 2 Enter...

Страница 16: ...TL2 A Com Port You can find the Com Port that was installed by clicking on start choosing Settings choosing Control Panel and then opening the System application Click the Hardware tab and then click...

Страница 17: ...note the COM number You can use this COM port number to connect to Hyperterminal or other serial communications programs T Th he er rm mo oP Pr ro ob be e I In nc c 112A JETPORT DR PEARL MS 39208 Tel...

Отзывы: