All refrigerated equipment generates heat as part of the normal operation. It is vital for correct performance and reliability
that adequate ventilation is provided to allow this heat to dissipate.
Any refrigerated equipment that is operated in a small unventilated area, especially in warm weather, can
cause the compressor to overheat and may result in premature failure.
NOTE: Failure to observe ventilation guidelines will void the manufacturer’s warranty!
All TELR/TLR Refrigerators require ventilation around them. 100mm (10cm) on each side and the back is re-
quired. See diagrams below showing correct placement and ventilation distances.
LOCATION & INSTALLATION
I
Do not store items on top of the fridge as this will also aff ect ventilation!
Level:
The cabinet should be standing on level ground to ensure correct operation of the self-closing door, and proper draining of
condensate from the evaporator (internal cooling surface).
Castors:
Where lockable castors are fi tted they must be locked to prevent accidental cabinet movement. This can be achieved by press-
ing the lock-down tab with your foot. Ensure that both castors are locked.
Mains connection:
This cabinet is suitable for connection to a standard 10Amp, 240 volt, 50Hz, supply. A dedicated outlet should be used for the
supply, do not use power boards or the like. A 3-pin moulded plug is fi tted as standard to the mains lead.