5.6.5
Flow Connections to the Flow Cell
Connect the inlet capillary and waste line to the flow cell when the flow
cell is installed to the device.
NOTICE
Backpressures that exceed the specified maximum pressure limit of the
flow cell can destroy the flow cell. Observe the following:
•
Use only the waste line connection that is provided for your
detector.
•
Connect the waste line to the flow cell only as described in the
manual.
•
Do not discharge waste from the flow cell through the open leakage
drain system of the Vanquish system.
•
Never expose the flow cell to excessive backpressure.
•
Avoid clogging of the flow cell or waste line.
•
When connecting a component in the flow path after the flow cell,
observe the specified backpressure for the flow cell.
In addition, observe the guidelines for proper connection of the flow cell
in
General Information and Guidelines (
.
Parts required
•
Inlet capillary, depending on your system arrangement:
¨
The detector is the only detector in the Vanquish system
Use the inlet capillary from the system ship kit.
¨
The detector is the second detector in the Vanquish system
Use the larger ID capillary shipped with the detector.
•
Detector waste line
For instructions on connecting the waste line, follow the steps in
Connecting the Detector Waste Line (
.
Tools required
For the detector waste line:
Tubing cutter (optional)
Preparations
1. Flush the system modules and capillaries upstream of the device to
waste before you connect the flow cell to the system flow path.
Refer to the
Vanquish System Operating Manual
.
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...