5
•
Installation
Page 64
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
5.6
Setting Up the Flow Connections
5.6.1
General Information and Guidelines
When setting up flow connections, follow these rules and
recommendations:
Flow connections can be filled with hazardous substances. Observe the
warning messages and precautionary statements presented in
.
NOTICE
Particulate matter from other system modules and components can
deposit in the flow cell and clog it.
•
Before you connect the flow cell to the flow path, make sure that
you thoroughly flush the modules in the system flow path upstream
of the device to waste.
•
When you install devices or components to the system, always flush
them to waste before connecting them in the system flow path. To
flush the Vanquish modules, follow the instructions in the
Vanquish
System Operating Manual
.
NOTICE
Flow cells are highly sensitive to contamination, clogging and high
backpressures. Even if the pressure exceeds the upper limit for a very
short time only, the flow cell may be permanently damaged. Observe
the following notes when connecting the flow cell to the system flow
path:
•
When connecting a component in the flow path after the flow cell,
observe the specified backpressure for the flow cell.
•
Use only clean Viper capillaries which were provided for the flow cell
and which have been properly protected by their cap before.
•
Use only the waste line which was provided for the flow cell.
•
Avoid clogging of the flow cell or waste line.
•
Improperly set up flow connections can lead to leaks on the flow cell
or even destroy the flow cell. Do not invert the flow cell inlet and
outlet.
Содержание Vanquish VC-D50
Страница 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 154: ...7 Maintenance and Service Page 154 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Страница 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...
Страница 188: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...