Working Area Requirements
Location
Thermo Scientific
iCE 3000 Series Pre-Installation Requirements Guide (P/N 9499 500 30031, Version 1.6)
5-3
Location
The instruments are designed for use on a normal laboratory bench. See
Figure 5-1.
Typical location diagram for iCE 3000 Series AA spectrometers
printer
PC
mouse pad
Furnace
420 mm
480 mm
222 mm
780 mm
282 mm
384 mm
784 mm
480 mm
Ensure that the exhaust is correctly positioned above the LH
compartment (flame, EC100, T-Cell or Mercury-Cell) and RH
compartment (furnace). Make sure that the ventilation is on before
igniting the flame, using the EC100, T-Cell or Mercury-Cell or before
firing the furnace. Do not run any of those when the ventilation is
turned off.
Organic or volatile solvents should not be stored, even for a short time,
near the instrument.
Tip
Do not position the equipment so that it is difficult to operate the
extraction, electrical supply, cooling water, purge gas and plasma gas
controls.