Third-party software
Before installing third-party software on the computer running the desktop software,
confirm that the third-party software will not do the following:
• Restrict Ethernet communication.
• Interfere with instrument or computer operation.
Network connection options
You can connect an instrument to a network or computer in the configurations listed
below. For more information, see Appendix B, “Connect the instrument to a
network”.
Direct connection
Local area network (LAN)
connection
Connect platform connection
wired
wired
or
wireless
wired
or
wireless
Types of runs
Standard curve run
A standard curve run determines the absolute target quantity in samples.
1. The software measures amplification of the target in a standard dilution series
and in test samples.
2. The software generates a standard curve using data from the standard dilution
series.
3. The software uses the standard curve to interpolate the absolute quantity of
target in the test samples.
Chapter 1
Product information
Types of runs
1
30
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide