The syringe tip must be in front of the array when filling the case with immersion fluid.
7.
While holding the case
vertically
, remove the syringe tip, insert the screw end of the OpenArray
™
plug into the port of the case, then rotate clockwise until the black handle breaks off.
IMPORTANT!
To avoid leaking of immersion fluid, hold the case
vertically
and rotate the plug
slowly.
If the plug handle breaks off prematurely, use a Phillips #0 screwdriver to complete this step.
8.
If needed, clean the case with a laboratory wipe that has been thoroughly sprayed with ethanol,
then dry the case with a clean laboratory wipe.
The plates are ready for PCR on a compatible Applied Biosystems
™
real-time PCR instrument.
Chapter 9
Prepare the OpenArray
™
Plate for PCR
Seal the OpenArray
™
Plate
9
62
QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System User Guide