4
Nucleic Acid Applications
Measure dsDNA, ssDNA or RNA
20 NanoDrop Eight User Guide
Thermo Scientific
dsDNA, ssDNA and RNA reported values
• created on (date sample measurement was taken)
Settings for Nucleic Acid Measurements
Baseline Correction
For the dsDNA, ssDNA or RNA, enter baseline correction wavelength in nm or use
default value (340 nm)
Optional user-defined baseline correction
. Can be used to correct for any
offset caused by light scattering particulates by subtracting measured absorbance at
specified baseline correction wavelength from absorbance values at all wavelengths
in sample spectrum. As a result, absorbance of sample spectrum is zero at specified
baseline correction wavelength.