iSort
™
Automated Cell Sorter User Guide
35
Load sample
1.
When prompted by the
Insert Sample
dialog (page 34), open the sort
compartment door to access the sample input port and the collection area.
2.
Make sure that the sample tube holder is in the down position and the
collection tray is pulled out.
3.
Place the 12 × 75-mm tube containing your sample into the sample tube holder.
Note:
The sample tube must contain at least 250 µL of sample volume to
begin Analysis or Sort. For guidelines on sample preparation, see page 32.
4.
If sorting, insert the desired collection tubes into Position 1 and Position 2 of
the collection tube holder. The collection tube holder can hold two 1.5-mL,
5-mL (12 × 75-mm), or 50-mL tubes in any combination.
•
If you wish to sort into 1.5-mL or 5-mL tubes, insert a tube adapter into the
desired position on the collection tray to hold your collection tubes.
•
If you wish to sort into 50-mL tubes, remove the tube adapter and insert the
50-mL tube directly into the desired position on the collection tray.