iSort
™
Automated Cell Sorter User Guide
33
System preparation
Before starting
•
Make sure that the collection tray is pulled out and there are no collection tubes
present.
•
Place an empty 12 × 75-mm tube into the sample tube holder in the down
position.
Choose application
1.
On the
Home Screen
, press
Sort
or
Analyze
to select the corresponding
application.
2.
On the first screen of the selected application, choose
Cells
or
Particles
.
•
Select
Cells
for cells that express green fluorescent proteins (GFP), are
labeled with green extra cellular labels (FITC or AlexaFluor
™
488 dye), or
have been stained with green dyes (e.g., Sytox
™
Green dye)
•
Select
Particles
for fluorescent beads (e.g., Attune
™
Performance Tracking
Beads)
3.
When prompted, enter your name in the
Operator Name
text box using the
alpha-numeric keyboard, then press
OK
.
The name you enter will appear as the
File Owner
next to the name and size of
the saved FCS file in the
File Viewer
screen (see “Manage files”, page 46).
4.
After you have entered the operator name, the system automatically initiates
the
System Preparation
(page 34).