Herasafe 2030i / Maxisafe 2030i
50155215 |
133
Maintenance |
Chapter 10
Exchanging SmartPorts
1. Before replacing the SmartPorts, open the front window. Doing so will produce an
alarm to indicate that there is a potentially unsafe operating condition.
2. Before replacing the SmartPorts, disinfect the workspace, front window, as well as
the inner and outer compartment walls.
3. First remove the inner grommet of the SmartPort in the workspace inwards from the
side panel opening.
4. Disinfect the grommet in the workspace and disinfect the inner compartment walls.
5. Insert new inner grommet in the side panel opening from the inside.
6. Then remove the outer grommet of the SmartPort in the workspace inwards from the
side panel opening.
7. Place the removed grommet in the workspace and disinfect.
8. Disinfect the outer compartment walls.
9. Insert new outer grommet in the side panel opening from the outside.
10. Safely discard used disinfected grommets.
Retrofitting and Repairs
External communication systems, e.g. failure report systems or components for supplying
media such as gas solenoid valves can be retrofitted and integrated into the device control.
Contamination!
There is a risk of contamination for users and the product itself. In order to
prevent contaminated particles escaping from the workspace, the unit must be
switched on in order to exchange the SmartPorts.
The SmartPorts are removed inwards into the workspace. Prior to replacement,
all work in the workspace must be finished.
Repairs
All retrofitting and repair work are interferences with the safety system of the
unit. Particularly modifications to the filter system and resulting changes of the
airflow may impair personal and material protection. Such work must be carried
out only by authorized service personnel.
Содержание HERASAFE 2030i Series
Страница 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Страница 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Страница 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Страница 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Страница 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Страница 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Страница 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Страница 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Страница 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Страница 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Страница 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...