Operating the Microscope
Thermo Scientific
DXRxi User Guide
29
Using Optical Polarizers
Optional polarizers are available for the
and the
. There are two fixed polarizers that send polarized light to the sample (one for
each type of illuminator).
Two
are also available (configurable and fixed). The polarization
analyzers can be used to capture any change in polarization that is due to the state of the light
interacting with the sample. The configurable polarization analyzer can be used with both of
the fixed polarizers (reflected light and transmitted light). The fixed polarization analyzer
works only with the fixed reflected light polarizer.
Figure 27.
Optional fixed polarizers (send polarized light to the sample)
Figure 28.
Optional polarization analyzers (can be used to capture any change in polarization that
is due to the state of the light interacting with the sample)
NOTICE
Store polarizers in shipping cartons when not in use. Polarizer
performance may deteriorate with extended exposure to light.
Fixed transmitted light polarizer
Fixed reflected light polarizer
Fixed polarization analyzer
Configurable polarization analyzer
Содержание DXR xi
Страница 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Страница 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Страница 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Страница 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Страница 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Страница 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Страница 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Страница 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...