3.
Press
ENTER
. The
Internal Printer Test
screen appears.
4.
Press
Start Test
to run the test. The test print
routine appears on the printer.
While the test is running, you can press
Stop
Test
to stop it.
If the test fails, follow these guidelines:
•
Make sure that the internal printer is the
selected printer device on the
Utilities
screen.
•
Make sure that the internal printer is installed
correctly. (Return to the main screen and press
ENTER
. If the paper does not move, the printer
may not be installed correctly.)
•
Make sure that the thermal paper is threaded
with the thermal side toward the printer head
(the outside surface of the roll is the thermal
surface).
•
Contact technical support for more
troubleshooting advice.
RS232C Test
The RS232C test allows you to verify that the
RS232C port is functional. To run the test, you will
need to have the RS232 Test Plug (336035)
attached to the RS232C port on the back panel of
the instrument. Running the RS232C test takes
approximately five minutes.
1.
Verify that the RS232 Test Plug is securely
attached to the RS232C port on the back panel
of the instrument.
2.
With the
Performance Validation
screen
displayed, use the arrow keys to highlight
RS232C Test
.
3.
Press
ENTER
. The
RS232C Test
screen
appears.
4.
Press
Start Test
to run the test. When the test
is complete, the message "RS232 OK" or
RS232 Failed" appears on the screen.
While the test is running, you can press
Stop
Test
to stop it.
If the test fails, follow these guidelines:
•
Make sure that the test plug is installed
correctly and is not loose or damaged.
•
Contact technical support for more
troubleshooting advice.
Analog Output Test
The Analog Output test sends an output signal to
an analog recorder. The signal alternately ramps
from low to high voltage and back again. This
produces a plot on the recorder that will vary
depending on the paper speed. To run the Analog
Output test, you will need to have an analog
recorder connected to the instrument.
1.
Make sure that the analog recorder is securely
connected to the instrument.
2.
With the
Performance Validation
screen
displayed, use the arrow keys to highlight
Analog Output Test
.
3.
Press
Enter
. The Analog Output Test screen
appears.
4.
Press
Start Test
to run the test. While the test
is running, you can also press
Stop Test
to
stop it.
If the test fails, contact your distributor for technical
support.
Using the Performance Validation Program
3-6
GENESYS 10 Operator’s Manual
Содержание GENESYS 10
Страница 1: ...GENESYS 10 Series Spectrophotometers Operator s Manual ...
Страница 2: ...Copyright 2002 Thermo Spectronic All rights reserved ...
Страница 8: ...GENESYS 10 Operator s Manual ...
Страница 9: ......
Страница 10: ......
Страница 14: ...iv GENESYS 10 Operator s Manual ...
Страница 18: ...Table of Contents viii GENESYS 10 Operator s Manual ...
Страница 50: ...Using the Instrument 2 28 GENESYS 10 Operator s Manual ...
Страница 60: ...Connecting Using Accessories 4 4 GENESYS 10 Operator s Manual ...
Страница 64: ...Performing Maintenance Procedures 5 4 GENESYS 10 Operator s Manual ...
Страница 72: ...Appendix C Calculations C 4 GENESYS 10 Operator s Manual ...
Страница 76: ...Index I 4 GENESYS 10 Operator s Manual ...