5
•
Installation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 91
2. Connect the delay capillary between the detector flow cell outlet or
the outlet (OUT) on the flow cell connection unit and the fraction
collection valve central port.
Figure 37: Connecting delay capillary to central port of fraction collection
valve
3. Route the delay capillary through the recess in the partition panel of
the detector.
4. Route the delay capillary between detector and the fraction
collector depending on length of the delay capillary that is used.
5.9.4
Connecting the Flush Buffer Loop
Parts required
Needle capillary
TIP
Make sure to check the flow rate of application and choose the
proper needle capillary according to application flow rate range.
Meanwhile, the installed flush buffer loop must be compatible with
specific needle capillary. See below table for details.
Flow rate
Needle Capillary
ID
Flush Buffer Loop
Volume
Remark
0.05
–
5 mL/min
0.18 mm
50 µL
Included in ship kit
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...