7
•
Maintenance and Service
Page 172
Integral Fraction Collector FT (VF-F20-A) Operating Manual
7.9
Replacing the Needle Capillary
When
If the needle capillary is clogged or damaged, or a new needle capillary
needs to be installed due to flow rate change.
Parts required
Replacement needle capillary
Figure 71: Needle capillary
No.
Description
No.
Description
1
Needle capillary Viper fitting
on the needle unit
5
S-shape slot on the foam part
2
Needle capillary screw
6
Needle capillary press plate
3
Needle capillary support block
7
Needle capillary Viper fitting on
the fraction collection valve
4
Needle capillary
Tools required
Screwdriver, Torx T10
Preparations
1. Prepare the device for maintenance, move the needle unit into
service position. See
page 153).
2. Turn off the device with its main power switch.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...