5
•
Installation
Page 66
Vanquish Pumps (VH-P10-A-02) Operating Manual
c) Screw the filter top to the filter bottom.
Figure 14: Assembling the solvent line filter
•
On the pump inlet, remove the plug that protects the inlet during
shipment, if applicable.
Follow these steps
1. Connect each solvent line to the pump inlet port for which it is
intended (see the labeling on the solvent lines and in the pump).
TIP
For easy access to the lower degas chambers, remove the seal wash
detector assembly (funnel and detector). Grasp and pull the assembly
upward by the funnel. The detector including the tubing and cable can
remain connected to the funnel.
Figure 15: Removing the seal wash detector assembly
2. Route the solvent lines from the pump inlet to the solvent rack,
through the tubing guides in the pump and all modules above the
pump in the system stack. Observe the routing guidelines in the
Guiding Capillaries and Tubing Through the System
section in this
manual.
No.
Description
1
Solvent line filter, bottom
2
Filter frit
3
Solvent line filter, top
No.
Description
1
Funnel, seal wash detector
2
Seal wash detector
3
Detector cable
4
Seal wash line
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...