3
•
Pump Overview
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 33
3.2
Operating Principle
The pump can generate pressures up to 151 MPa with ultra-low
pulsation. The pump is based on a two-channel, dual-piston design with
electronic compressibility compensation.
The picture illustrates how the pump operates:
Figure 1: Operating principle
No.
Description
1
Solvent selector (pump inlet)
2
Solvent degasser (vacuum degasser)
3
Pump head, left (two pistons, parallel)
4
Y-connector, left
5
Pump head, right (two pistons, parallel)
6
Y-connector, right
7
Purge unit
8
System pressure sensor
9
Inline filter (static filter)
10
Pump outlet (outlet of the inline filter)