Test Procedures
Page 42 of 60
M1393R0
OQ/PQ Operating Instructions
FIGURE 4-15 The Properties Report Column Box
f)
Check that the segment that is used for calculation is in the middle of the step.
g)
Modify the values in the ‘Formula’ field (FIGURE 4-15). Each segment used for the
calculation is defined by a start and a stop time (e.g. in the example in FIGURE 4-15
starts at 8 minutes and ends at 9 minutes).
h)
Check that the test result is correct and save the report.
4.14.2
Completing the Cover Page and Printing
Some of the instrument parameters required for the report can not be completed
automatically by CHROMELEON
and require manual input (e.g. customer information, S/N
of the FAMOS Microautosampler).
To complete the ‘Basic Tests’ page of the report:
a)
Open the report (PQ_OQ_LCP.rdf).
b)
Choose the ‘Specifications’ sheet.
c)
Select the ‘Instruments’ section and enter the missing information (e.g. S/N of the
FAMOS, flow cell, thermometer, etc.).
d)
Select the ‘Additional Information’ section and enter the customer information (e.g.
name, operator, position, etc.).
e)
Choose the ‘Other Tests’ sheet and enter the result of the manual tests (e.g. Section
4.2, 4.3, 4.4, etc.).
f)
Check the result of the Gradient Accuracy Test as described in Section 4.14.1.
g)
Save the report.
h)
Print the report using the ‘Batch Report’ option form the ‘File’ menu.