Process
OQ/PQ Operating Instructions
M1393R0
Page 15 of 60
3.2
Preparations
The system components that are included in the system configuration have to be prepared
according to the following steps before starting the OQ/PQ procedure.
3.2.1
CHROMELEON® Setup
All CHROMELEON programs required to perform the OQ/PQ test procedures are provided
as a CHROMELEON backup file on the CD ROM ‘IQOQPQ on UltiMate
™
(Plus) Systems’
(P/N 163935). Two folders are provided, one is for the Nano LC (NAN) and capillary LC
(CAP) configurations. In addition, different CHROMELEON server configuration files are
available.
It is assumed that the service engineer has basic knowledge of the UltiMate system and
CHROMELEON software. Please refer for more detailed information about the installation
and usage of the UltiMate system and CHROMELEON to the manuals provided along with
the products. A detailed description of the tests is provided in Chapter 4.
3.2.1 A
Restoring the Backup Files
Refer to the CHROMELEON on-line help (F1 key) for more information on how to restore
backup files.
To prepare the PC and the CHROMELEON software for the OQ/PQ test procedures:
a)
Restore the CHROMELEON backup file from the CD ROM. All programs and sequences
are now available for the test procedures (FIGURE 3-1).
FIGURE 3-1
OQ/PQ Programs after Restoring
b)
Copy the three server configuration files that are available on the same CD ROM into
the
Chromel/Bin
directory of the computer.