Страница 1: ...Cryogenic Storage Freezers Installation and Operation 325440H01 Rev A December 2013 Visit us online to register your warranty www thermoscientific com labwarranty...
Страница 2: ...med by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific ma...
Страница 3: ...specifications in this document supersede all previous information received by the purchaser Thermo Fisher Scientific Inc makes no representations that this document is complete accurate or error free...
Страница 4: ...n 6 Wiring 6 Leveling 7 Operation 8 Control Panel Features 8 Initial Loading 10 Start Up 11 Frame Heater Switch 14 Backup System Optional 15 Maintenance and Troubleshooting 19 Condenser Maintenance 19...
Страница 5: ...orage Freezers Installation and Operation Manual 1 1 Cryogenic Models The following cryogenic models are described in this manual Model Number Suffix ULT 10140 9 D 19 ULT 10140 9 M 24 ULT 10140 9 V 23...
Страница 6: ...avoided could result in serious injury or death WARNING This symbol indicates situations where dangerous voltages exist and potential for electrical shock is present The snowflake symbol indicates ext...
Страница 7: ...nner not specified by the manufacturer the protection provided by the equipment may be impaired Do not modify system components especially the controller Use OEM exact replacement equipment or parts B...
Страница 8: ...time period This activity could also cause damage to the cabinet WARNING DO NOT store flammable liquids in this unit Avoid opening the lid for extended time periods since chamber temperature air will...
Страница 9: ...relative humidity 60 for temperatures up to 32 C 90 F Main supply voltage fluctuations not to exceed 10 of the nominal voltage 4 1 Electrical Specifications The last character in the model number lis...
Страница 10: ...open at least 90 degrees Do not position the equipment in direct sunlight or near heating diffusers radiators or other sources of heat The ambient temperature range at the location must be 59 to 77 F...
Страница 11: ...the grounding prong from the service cord plug If the prong is removed the warranty is invalidated 5 3 Leveling The unit must be level both front to back and side to side To level chest models leave t...
Страница 12: ...Power On Indicator This indicator lights when power is connected to the freezer 3 Temperature Failure Indicator This indicator lights when the freezer temperature deviates either above or below the a...
Страница 13: ...showing the cabinet temperature 13 Digital Temperature Display Window This window displays chamber temperature alarm values etc depending on the operating status of the freezer and the procedure being...
Страница 14: ...stem Battery Low Indicator 6 2 Initial Loading After reading and completing the Safety Considerations Pre Inst0allation Installation and Operation sections of this manual turn the Key Switch to the PO...
Страница 15: ...the Setpoint Security will be deactivated in this key position 3 The unit will start about 30 seconds later 6 3 2 Setting the Cabinet Temperature To set the cabinet temperature complete the following...
Страница 16: ...int Security disabled 2 Press and hold the Warm Alarm Setpoint pad The LED next to this pad lights The temperature display shows the Warm Alarm value 3 Press and hold or to adjust the Warm Alarm Setpo...
Страница 17: ...will decrease to approximately 15 times per minute The indicator will then remain flashing at this rate unless the condition reoccurs until the Alarm Reset button is pressed Once this has occurred th...
Страница 18: ...n silence the alarm in this manner as many times as necessary until the sensor cools below the warm alarm temperature The alarm quits automatically when the sensor cools below the warm alarm value Not...
Страница 19: ...re 3 Note Always purchase 22 psig supply cylinders and ensure that they are equipped with siphon tubes for withdrawing liquid from the bottom of the cylinder LN2 bottles are functional at any reasonab...
Страница 20: ...re that the injection solenoid is energized to allow all the liquid to bleed off instead of being trapped in the supply tubing Failure to do this results in activation of the pressure relief device wh...
Страница 21: ...40 C the backup system setpoint should be no colder than 135 C Note The temperature shown in the display window is accurate to 1 C The backup system can run for a minimum of eight hours on a fully cha...
Страница 22: ...2 375 LN2 Inlet Conn Copper Tubing Factory Supplied 0 625 O D MACHINE COMPARTMENT REAR VIEW Table 1 1 Backup System Flow Rates Cryogenic Freezers Lbs hr 75 F Ambient Temperature Flow Rate Center Air T...
Страница 23: ...he following steps 1 Pull the grill open 2 Remove the filter Check the fans If a fan is not operating contact an Authorized Service Company immediately 3 Vacuum the condenser using a soft brush attach...
Страница 24: ...there is freezer odor wash the interior with a solution of baking soda and warm water Clean the exterior with any common household cleaner 7 Reinstall the sub lids close the lid restart the freezer a...
Страница 25: ...leshooting Procedures Problem Details Check Fuse indicator is on Surge protector is blown Clean Filter indicator is on Condenser filter is dirty Refer to Section 7 1 on page 19 Alarm Battery Low indic...
Страница 26: ...2 Connect the nine volt DC battery located at the recorder s upper right corner This battery provides backup power 3 Install clean chart paper refer to Section 8 3 below 4 Remove the plastic cap from...
Страница 27: ...old the Change Chart button 3 in Figure 6 for one second The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and tim...
Страница 28: ...that the program is correct 1 Press button 3 in Figure 6 and allow the pin to move off the scale 2 Press 1 for 10 seconds and release 3 Confirm that the green LED flashes 3 times followed by a pause I...
Страница 29: ...ts light bulbs and door gaskets are excluded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to th...
Страница 30: ...eing performed Expendable items i e glass filters pilot lights light bulbs and door gaskets are excluded from this warranty Replacement or repair of component parts or equipment under this warranty sh...