Tuning and Calibrating
Creating a Quality Control (QC) Tune Report
72
TSQ Quantum XLS Series User Guide
Thermo Scientific
Figure 40.
QC tune report
Y
To create and save a QC tune report
1. Choose
View > QC Report
to open the TSQ QC Report dialog box.
2. Click
Start
to start the QC report procedure.
3. After the procedure terminates, click
View Report
to view the QC report as a PDF file.
4. Save the PDF file.
Q Q
Q
p
Dec 02, 2011 15:51:56 TSQ Quantum XLS Ultra S/N: TQU01792
Device SetPoint Readback
Lens 1 (V) -45.0 -45.5
Lens 2 (V) -180.0 -180.7
Lens 3 (V) -25.0 -25.2
Lens 4 (V) -10.0 -10.0
Lens 1-1 Offset (V) -9.6 -10.0
Emission Current (uA) 50 48
Electron Lens (V) 10 10
Electron Energy (eV) -70 -69
Source Temperature (C) 100 100
Ion Gauge Pressure (Torr) 7.1e-06
Q1 Peak Width(amu): 0.7
Air/Water Leak Check: FAIL
qcr20111202155106
#
1
ST:
3.02
uS:
1
NL:
2.07E7
68
70
m/z
218
220
m/z
500
502
504
m/z
0
10
20
30
40
50
60
70
80
90
100
R
el
ati
ve Intens
ity
0
10
20
30
40
50
60
70
80
90
100
R
el
ati
ve Intens
ity
0
10
20
30
40
50
60
70
80
90
100
R
el
ati
ve Intens
ity
71.213
69.030
219.021
501.865
qcr20111202155106
#
11
ST:
1.02
uS:
1
NL:
2.76E7
+ c CI Q1MS [46.500-700.500]
100
200
300
400
500
600
m/z
0
20
40
60
80
100
Rel
at
ive I
nt
ensi
ty
218.977
130.966
263.972
501.946
413.943
99.960
463.946
313.982
613.997
375.932
551.963
673.685
Cal. Mass(m/z) Exact Mass(m/z) Intensity(cts) Ion Ratio(%)
68.971 68.995 14484893 52
218.977 218.986 27647784 100
501.946 501.971 6361739 23
Signature_____________________________ Date________________________
Содержание TSQ Quantum XLS
Страница 1: ...TSQ Series TSQ Quantum XLS Series User Guide 70111 97225 Revision A December 2011 ...
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 32: ......
Страница 64: ......
Страница 76: ......
Страница 102: ...86 TSQ Quantum XLS Series User Guide Thermo Scientific Figure 50 Ion source assembly assembled and exploded views ...
Страница 160: ......
Страница 166: ......
Страница 176: ......