8
Using the Direct Sample Probe
Running the Sequence
154
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Figure 70.
Insert Probe message
2. Inset the probe into the inlet valve.
a. Insert the guide ball on the DIP probe into the guide ball hole on the guide bar.
b. Slide the probe forward into the inlet valve until the guide ball is at the guide bar’s
first stop. See
.
c.
Tighten the inlet valve knob clockwise to ensure a leak-tight seal.
Figure 71.
Probe at the first stop on guide bar
3. Click
OK
. The forepump evacuates the inlet valve.
4. When the Safe to Open Valve message to appears, click
OK
.