15
Calibrating the MS
Calibrating and Tuning the MS
64
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
Calibration Solution Peak Values
The mass spectra of the EMRS in positive and negative ion polarity modes (
, respectively) have peaks at
m/z
values close to the following theoretical values.
Figure 32.
Mass spectrum of EMRS in positive ion mode
Positive mode (
m/z
)
Negative mode (
m/z
)
69
69
102
113
215
302
622
602
922
1034
1522
1634
IMPORTANT
If you observe interfering peaks in the EMRS spectrum that are within
±10 Da of any of these calibration masses, follow the procedure
. Ensure that the interfering masses show less than 25 percent
of the intensity of the calibrant ions.
2