Recommended Inlet and Flow Rates
Use the following tables for the recommended settings when operating your system in H-ESI,
APCI, or NSI mode.
Guidelines for LC/H-ESI/MS Parameters
Guidelines for LC/APCI/MS Parameters
LC flow rate
(μL/min)
Spray
voltage
(V)
Gas (arbitrary units)
Ion transfer tube
temp (°C)
Vaporizer
temp (°C)
Nitrogen use
(L/min)
Sheath
Auxiliary
Sweep
Up to 15
Pos: 3500
Neg:
-2500
5
2
0
Altis: 325
Quantis: 275
Fortis: 275
20
3.5
16–99
25
5
0
75
8.0
100–199
35
7
0
Altis: 325
Quantis: 300
Fortis: 300
275
10.5
200–500
50
10
1
Altis: 325
Quantis: 300
Fortis: 300
350
15.0
LC flow
rate
(μL/min)
Sheath gas
arbitrary
units)
Auxiliary
gas
(arb
units)
Sweep gas
(arb units)
Ion transfer tube
temp (°C)
Vaporizer
temp (°C)
Corona
discharge
current (μA)
Up to 5
15
5
0
Altis: 300
Quantis: 250
Fortis: 250
275
Pos: 4
Neg: –10
5–199
25
325
200–500
45
Altis: 325
Quantis: 275
Fortis: 275
350
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...