Maintaining the MS on a Daily Basis
There are tasks that you can do before and after you operate the MS to prevent needless
downtime.
• Before Operating the Mass Spectrometer
• After Operating the Mass Spectrometer
Before Operating the Mass Spectrometer
These tasks ensure that the MS is ready for operation.
The tasks include:
•
•
Checking the Vacuum Pressure Levels
•
Checking the Argon and Nitrogen Gas Supplies
Finding and Fixing Air Leaks
A major air leak can indicate insufficient pressure levels to turn on the system. In the Tune
window, a green square,(
), indicates that the readback value is good. Possible causes of a
major leak might be a loose or disconnected fitting, an improperly positioned O-ring, or an
open valve.
1. Listen for a rush of air or a hissing sound coming from the MS.
• If you do not hear these sounds, there is no air leak.
• If you hear these sounds continue with step 2.
2. Follow
“Shutting the MS System Down Completely.”
3. Visually inspect the vacuum system and vacuum lines for leaks.
Note
You do not need to calibrate and tune the MS as part of your daily routine.
Generally, you must calibrate the MS every one to three months of operation for optimum
performance over the entire mass range of the mass detector.
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...