Setting Up Sample Introduction Techniques
Setting Up H-ESI or Low Flow H-ESI Direct Infusion
Thermo Scientific
TSQ Altis, Quantis, and Fortis Hardware Manual
135
Setting Up Low Flow H-ESI Direct Infusion
1. Place sample in the syringe (1
). The syringe pump is not shown.
2. Insert the grounding union (2) into the grounding union holder (3).
3. Use Viper fittings to connect the natural colored PEEK tubing (5) to the grounding
union (2) and to the H-ESI spray insert (4).
4. Use PEEK fingertight fittings to connect a length of red PEEK tubing (7) to the syringe
union adapter (1) and the grounding union (2).
No.
Description
1.
Syringe union
2.
Grounding union
3.
Grounding union holder
4.
H-ESI spray insert
5.
Red PEEK tubing connecting grounding union to the H-ESI source
6.
Red PEEK tubing connecting the syringe to the grounding union
Note
Use Viper fittings on the natural colored PEEK tubing (the nano-Viper capillary.)S
The other fitting should be PEEK.