Maintaining the API Source Interface
Removing the Ion Sweep Cone and the Ion Transfer Tube
Thermo Scientific
TSQ Altis, Quantis, and Fortis Hardware Manual
105
5. After the source cools to room temperature, remove the source housing, refer to the
OptaMax NG Ion Source User Guide
, Chapter 2 section
Removing and Installing the API
Source.
6. Remove the ion sweep cone by grasping its outer ridges and pulling it off (
necessary, loosen the screws on the ion sweep cone.
7. Align the flat edges of the custom removal tool with the flat edges on the exposed tip of
the ion transfer tube, and then do one of the following:
• TSQ Altis MS: Rotate the tool (1 in
) counterclockwise. When the tube is
free of the spray cone, use the hook (flat edges) on the tool (2) to slowly pull it out of
the API source interface.
Figure 54.
Ion transfer tube removal tool (TSQ Altis MS)
CAUTION
• Make sure that you do not accidentally lift the release lever that is located above
the API source interface, which will vent the MS.
• To avoid contaminating the ion transfer tube, do not touch its exposed entrance.
No.
Description
1.
Use this end of the tool to rotate the tube.
2.
Use this end of the tool to hook onto the tube and pull it out.
1
2
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...