3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
324
TSQ 9000 Hardware Manual
Thermo Scientific
.
6. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the board to the chassis. Loosening the screw will separate the
board from the instrument. The screw is captured, so do not attempt to completely
remove the screw from the board and bracket.
ATTENTION DOMMAGES À L’INSTRUMENT
Limitez au maximum la manipulation des
cartes. Ne touchez aucun composant de carte. Gardez la carte uniquement près du support
en tôle.
Figure 345.
Locating the Lens Driver Board
Tip
The bracket has a white diagram describing the location of this screw. It is
approximately two inches from the top of the board directly under the white label.
Lens Driver
Board
Bracket
Содержание TSQ 9000
Страница 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...
Страница 8: ......
Страница 124: ...1 Performing Routine Maintenance Cleaning the Filters 96 TSQ 9000 Hardware Manual Thermo Scientific ...
Страница 476: ......