Replacing the Chassis Feet
To replace the chassis feet on the bottom of the TSQ 9000 instrument
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the TSQ 9000 Instrument”
3. Follow the instructions in the section,
to remove the
components required to safely replace the chassis feet.
4. Remove the small and large glass covers in the manifold.
5. Carefully tip the TSQ 9000 instrument and rest it on its left side.
6. Use a T20 Torxhead screwdriver to remove the screws holding any damaged feet to the
bottom of the instrument.
7. Attach the new chassis feet.
8. Set the instrument upright.
9. Replace the large and small glass manifold covers.
10. Follow the instructions in the section,
to replace the
components you removed.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 318.
Removing the Feet on the Bottom of the TSQ 9000 Instrument