![Thermo Scientific TSQ 9000 Скачать руководство пользователя страница 365](http://html.mh-extra.com/html/thermo-scientific/tsq-9000/tsq-9000_hardware-manual_1104856365.webp)
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
337
Distribution Board
The distribution board lies on the bottom of the TSQ 9000 instrument chassis.
To replace the distribution board
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 361.
Locating the Left Hand Front Panel Screws
4. Lift the left hand front panel off the instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. If you are using
hydrogen as a carrier gas, power off the GC.
Screws
Left Hand Front
Panel
Screw on Front
of Instrument
CAUTION - INSTRUMENT DAMAGE
Limit handling of the boards as much as possible.
Do not touch any board components. Hold the board by the sheet metal bracket only.
Содержание TSQ 9000