5
Service Menu Item
Installation Item
Thermo Scientific
TriPlus RSH User Guide
207
Configure SPME Tool
Working with the SPME Tool requires two adjustments of the SPME Fiber.
Configure SPME Tool parameters are described in
The last wizard point was Teach Tool Station. For this task, the tools had to be removed from
the slots. After finishing the teaching task, the wizard will move to the next point,
Configure
SPME Tool
. If the SPME Tool has not been positioned in a slot at the moment of activating
this wizard point, the task will be grayed out.
To overcome the situation
1. Insert the SPME Tool manually in a slot.
2. Select the previous wizard point
Teach Tool Station
again.Wait until the first movements
of the Z-Axes are finished.
3. Press
Back
and leave this wizard point without going through all the steps.Check if the
wizard point Configure
SPME Tool
is active.
To configure the SPME Tool
1. Select the wizard item
Configure SPME Tool
. See
.
Figure 147.
Wizard Item Configure SPME Tool
2. When a tool other than the SPME Tool is installed in the TriPlus RSH Injection Unit,
the wizard asks for a tool change.
Table 92.
Configure SPME Tool Parameters
Item
Description
Fiber Alignment
Fiber position is aligned with the needle tip.
Standby Position Offset
Fiber is inserted 2 mm into sheath metal needle.
(2 mm retraction is a recommended value). When using very
soft septa, adjust to 3.0 mm.