TriPlus RSH User Guide
Thermo Scientific
Tray Types
A Tray Type contains all the information about the geometry of the Tray.
Parameters such as how many sample vials are in a column and in a row, and the distance from
vial to vial are all summarized in this class for each Tray Type.
Selecting the list item
Tray Types
opens up the following parameters. See
.
Adjust Needle Penetration
Adjusts the Needle Penetration Depths in the vial.
Manually turn the needle as far down in the vial as required.
The cut-out in the Tray Holder frame as well as in the Tray
allows visual control of the needle tip.
Note
: The value found is not stored automatically. This task
allows a check only. If the needle penetration must be
changed, select the Tray Type to enter the new established
Needle Penetration value.
Check Teaching
Checks the teaching position of the module.
This task serves as a quick check to determine if the X-, Y-,
and Z-positions are defined correctly.
In case a teaching point is incorrect, return to the Menu
Setup
for readjustment of the teaching position.
Note that this action can be done under the
Extended User
Level
but not under
User Level
.
Teach Module
The positions of the X-, Y-, and Z-axes can be defined again,
teaching.
Table 69.
Tray Holder Option pull-up Menu Parameters When Tray(#) is Selected (Sheet 2 of 2)
Item
Description