3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
135
Tools
In the class Tools are the various tool Types defined. Selecting the list item
Tools
opens up the
.
Figure 90.
Menu Item Tools and Menu List Items Tools
•
shows the Syringe Type selection for HS, LS, and SPME Tools.
•
shows the Syringe Type selection for ITEX Tool.
Delete MHE Station
Deletes of the present. MHE Station.
Check Teaching
Checks the teaching position of the station.
This task serves as a quick check to determine if the X-, Y-,
and Z-positions are defined correctly.
In case a teaching point is incorrect, return to the Menu
Setup
for readjustment of the teaching position.
Please note that this action can be done under the
Extended
User Level
but not under
User Level
.
Teach Module
The positions of the X-, Y-, and Z-axes can be defined again,
(teaching).
Reset to Default
Resets the selected module parameter to default value.
Table 65.
Tool Adapter Options Parameters (Sheet 2 of 2)
Item
Description