OC Actuator Options Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
.
Position Tolerance
Checks the difference between the taught X-Y-axes positions
against the actual position each time the head moves to the
OCI.
If the difference is greater than the
Position Tolerance
of
1.5 mm (default value) then the TriPlus RSH immediately
stops. This is a safety feature to avoid needle bending.
Table 43.
OC Actuator Parameters (Sheet 2 of 2)
Item
Description
Table 44.
OC Actuator Option Parameters
Item
Description
Reset to Default
Resets the selected module parameter to default value.
Check Teaching
Checks the teaching position of the module.This task serves
as a quick check to determine if the X-, Y-, and Z-positions
are defined correctly. In case a teaching point is incorrect,
return to the Menu
Setup
for readjustment of the teaching
position.
Please note that this action can be done under the
Extended
User Level
but not under
User Level
.
Teach Module
The positions of the X-, Y-, Z-axes can be defined again,
(teaching).
Check OC Alignment
Verifies the Needle Penetration in the OC Injector.
The head moves to the OCI teach position, checks the
Position Tolerance
, inserts the syringe needle and checks for
any resistance. If no resistance is felt the test is positive, the
needle can be inserted without bending.
Check OC Actuator
Verifies the opening and closing of the OC Actuator.