1
TriPlus 300 Headspace Site Preparation
Space and Load Requirements
Thermo Scientific
TriPlus 300 Headspace Preinstallation Requirements Guide
3
3. Allow at least 30 cm (12 in.) of clearance behind the system.
4. Make sure you have at least 90 cm (3 ft.) of clearance above the system.
5. Make sure your workbench can support a TriPlus 300 Headspace system. Remember,
additional instruments add to the total weight.
6. Ensure that your work area is stable and free of vibration from nearby equipment. The
system is a sensitive instrument.
GC-GC/MS TriPlus 300 Headspace Configuration Space and Mass
Requirements
TriPlus 300 Headspace sampling system must o be installed on the left of the GC-GC/MS
system. The maximum distance from the GC-GC/MS system must considers the length of
the transfer line that connects the TriPlus 300 Headspace sampling system to the GC injector.
Use
to determine the space and mass requirements according to your GC-GC/MS
TriPlus 300 Headspace sampling system configuration.
IMPORTANT
The measure reported in the tables are rounding-up for excess.
The tables do not include optional instruments e.g. computers, printers, etc. The
GC-GC/MS system should be placed on a workbench that has minimum dimensions of
0.75×2 m (2.5×6 ft.).
Table 2.
Space and Load Requirements
Instrument
Depth
Width
Height
Mass
cm
in.
cm
in.
cm
in.
kg
lbs
TriPlus 300 Headspace
Main Unit + Autosampler Unit
55
22
82
32
73
29
63
139
TRACE 1300
60
24
44
17
45
18
35
77
TRACE 1310
67
26
44
17
45
18
35
77
TSQ 8000 Series MS
89
35
40
16
45
18
61
135
ISQ Series MS
69
27
36
14
46
18
45
99
DSQ II MS
68
27
38
15
44
17
45
98
ITQ MS
68
27
38
15
44
17
45
98
Computer
1, 2
48
19
20
8
43
17
12
27
Monitor
2
16
7
46
18
32
13
4
8
Keyboard
2
23
9
46
18
50
2
1
2
1
This item is placed on the floor under the system, thereby reducing the weight requirements for your workbench.
2
Dimensions vary per manufacturer, therefore approximations are provided.