Chapter 10
Performing a Leaks Check (Capillary/Packed/UFM Columns)
Ensuring Tightness
Maintenance and Troubleshooting Manual
185
OPERATING SEQUENCE
Performing a Leaks Check (Capillary/Packed/UFM Columns)
This procedure allows to check for leaks on both injector and detector sides and it
is suitable for capillary, packed and UFM columns.
Before starting this sequence, install the column into the injector port and the
detector base body.
Materials needed:
•
Detector base body blind cap (for FID, NPD, FPD, PID, PDD)
•
6MB blind nut (for ECD)
•
Leak detector (Thermo Scientific GLD Pro, or equivalent)
1. In case the column is already installed, cool down the injector and the detector
base body to room temperature.
2.
Only for FID, NPD, FPD:
remove the detector. Do not remove ECD detector.
3.
Only for FPD:
remove the jet. In case of capillary columns, use typewriter
correction fluid to mark the position of the column entering the detector base
body. Lower the column end down to the top of the detector base body, then
tighten the M4 fixing nut to restore system tightness.
4. Finger tighten the relevant blind cap (fitted with a silicon seal) on the detector
base body, to seal the column flow path.
5.
Only for ECD:
remove the chimney and seal the line with the relevant blind
6MB nut.
6. Turn off any H
2
, air and make-up gas flows. Do not close the detectors gases
supplies.
7. Press
LEAK CHECK
to enter
Leak Check
menu.
LEAK CHECK COLUMN
Right column
1
Left column
1
<