10
Adding Systems
Adding an Auxiliary Gas System
534
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing and Connecting the Auxiliary Gas Module
To install and connect the Auxiliary Gas Module
1. Remove the cover of the external modules housing where installing the module. See
Figure 582.
Housing Cover Removal
a.
Using a T20 Torxhead screwdriver, unscrew and remove the left or right housing
cover screws.
b. Remove the cover from the housing.
2. Remove the manifold protecting plate
a.
Using a T20 Torxhead screwdriver, unscrew the two fixing screws, and remove the
protecting plate from the manifold. Save the protecting plate and the fixing screws.
Figure 583.
Manifold Protecting Plate
3. Install the module into the housing
Left/Right Housing Covers
Left Cover
Screws
Right Cover
Screws
Protecting Plate
Fixing Screws