Adding Systems
Adding the Oven Cryo System
502
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 538.
Aux Temperature/Cryo Module Installed into the GC
9. Install the solenoid valve.
The solenoid valve must be installed in the proper seat on the back of the GC. See
Figure 539.
Solenoid Valve Assembly Seat
a.
With care, take the solenoid valve assembly and remove the two screws from the
brackets using a T15 Torxhead screwdriver. These screws are used to fix the bracket
on the GC.
If not already done, connect the coolant tube to the solenoid valve using the proper
nut and ferrule. Use a 7/16-in. wrench for tightening the fittings. See