Adding Modules
Adding a FID, TCD/TCD In-Series, ECD, or FPD Detector Module
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
431
Adding a FID, TCD/TCD In-Series, ECD, or FPD Detector Module
This section provides instructions for adding a front/back FID, TCD/TCD In-Series, ECD,
or FPD detector module. According to the configuration of your TRACE 1300/TRACE
1310, a dummy module is present into the free site where the detector module is not installed.
See
Figure 446.
Add a Front/Back Detector Module
To add a new FID, TCD/TCD In-Series, ECD, or FPD detector module
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
3. Close the gas supplies.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
Dummy Back Detector Module
Dummy Front Detector Module
ATTENTION
Where a dummy module is installed, the gas connection is blocked by a
plug.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.