Installing Optional Kits
Installing the HS Adapter Kit on the SSL/SSLBKF Injector
344
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing the HS Adapter Kit on the SSL/SSLBKF Injector
This section provides the instructions for installing the HS Adapter kit PN 19050732, on the
SSL/SSLBKF injector for the use with a TriPlus 300 Headspace sampling system. See
Figure 345.
HS Adapter
To install the HS adapter
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
3. Close the gas supplies.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. Remove the column end from the injector and the detector.
a.
Open the front door of the GC.
Note
By pressing the
Maintenance
button, the GC cool down is automatically car-
ried out.
HS adapter