5
GC Main Frame Advanced Maintenance
Removing/Replacing the GC Top Cover
196
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Removing/Replacing the GC Top Cover
Perform this operation for accessing the internal compartments of the GC.
To remove/replace the top cover accessing the top parts of the GC
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
3. Power off the GC.
a.
Close the gas supplies.
b. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
c.
Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
4. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
5. Remove the autosampler if present.
6. Remove the cover.
a.
Using a T20 Torxhead screwdriver, unscrew the four screws that held it in place.
See
Figure 182.
Top Cover Removal (1)
WARNING
This operation must be carried out by authorized and trained Thermo Fisher
Scientific Service Field Engineers.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Top Cover Screws