Performing Injectors Routine Maintenance
Maintaining a Programmable Temperature Vaporizing Injector (PTV)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
123
13. If present, move the autosampler towards the module to restore the original alignment.
14. Turn the carrier gas on.
15. Set the normal injector, detector, and GC working conditions.
Replacing the PTV Broken Liner
To replace a broken liner
Figure 106.
PTV Injector: Broken Liner Replacement
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
Liner Seal
Liner
Silver Seal
Terminal Fitting
Bottom Fittings
Septum Cap
Septum
Liner Cap
Ferrule
Split Nut