![background image](http://html.mh-extra.com/html/thermo-scientific/thermoscientific-vanquish-uhplc-system/thermoscientific-vanquish-uhplc-system_operating-manual_1104838095.webp)
6
•
Operation
UHPLC System
Operating Manual
Page 95
6.6.4
Running Automated Sample Analysis
Automated sample analysis with the chromatography system involves
creating and running a sequence. In a sequence, you determine how the
injections of a sample are analyzed and in which order they are
processed. Each injection of a sequence is processed by using an
instrument method, which defines the property values and timed
control commands for each module. For details how to create an
instrument method using the Instrument Method Wizard, refer to the
Help
.
Chromeleon 7 supports two methods for creating a new sequence:
TIP
If no eWorkflows are available, you can create a new eWorkflow as
described in the
Help
. As an alternative, use the Sequence Wizard to
create the sequence.
•
eWorkflows (preferred method)
eWorkflows provide a set of predefined templates and rules for
creating a new sequence.
•
Sequence Wizard
Creating a Sequence via eWorkflows
1. On the
Console
, click the
eWorkflows
Category Bar.
2. In the
Navigation Pane
, click the eWorkflow name.
3. In the Work Area, click the instrument name, and then click
Launch
.
4. The eWorkflow Wizard guides you through the process. Complete
the steps in the wizard. For detailed information about a wizard
page, click the Help icon.
After you finish the wizard, the sequence is displayed in the Data
view of the
Console
.
Creating a Sequence via Sequence Wizard
1. On the
Console
menu bar, click
Create
.
2. The Sequence Wizard guides you through the process. Complete the
steps in the wizard. Specify the number of samples and/or
standards, the instrument method, processing method, and report
template that you want to use. For detailed information about a
wizard page, click the Help icon.
After you finish the wizard, the sequence is displayed in the Data
view of the
Console
.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...