5
•
Installation
UHPLC System
Operating Manual
Page 61
5.4.5
System Base Lock
A lock is integrated into the system base. Unlock the system base if you
want to move the system base and lock it if you want to prevent it from
being moved on the workbench. If your system configuration includes
one column compartment, you can switch between the moveable and
stationary mode using the system base locking tools shipped with the
system base.
Figure 17: System base locking tool
If your system includes two or three column compartments or a Charger,
a Thermo Fisher Scientific service engineer must dismount the
mentioned modules before you can (un)lock the system base with the
locking tools.
5.4.5.1
Unlocking the System Base
1. Have all but one column compartment and the Charger dismounted
by a Thermo Fisher Scientific service engineer.
2. Insert one locking tool into another.
Figure 18: Extended locking tool
3. Insert the combined locking tools into the lock port on the side of
the system base where the column compartment is installed.
If the column compartment is installed on the right side, insert the
combined locking tools into the right lock port as shown in the figure
below.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...