5
•
Installation
UHPLC System
Operating Manual
Page 53
5. Lift the column compartment into vertical position.
6. To insert the holding knobs in the grooves on the mounting bar,
grasp the column compartment by its sides and slightly lift the
column compartment up.
Figure 9: Attaching the column compartment to the system stack
5.4.1.3
Moving the System Stack after Installation
Moving the Vanquish system stack even over short distances on the
workbench can damage and impair the functionality of the system base.
Observe the following:
•
Configuration with one column compartment
:
Before moving the system stack, unlock the system base using the
locking tools. For details, see
.
•
Configuration with two or three column compartments or a Charger:
Before moving the system stack, have the mentioned modules
dismounted by a Thermo Fisher Scientific service engineer and then
unlock the system base. For details, see
.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...