7
•
Maintenance and Service
UHPLC System
Operating Manual
Page 109
7.5
Transporting or Shipping the System
NOTICE
Moving the Vanquish system stack even over short distances on the
workbench can damage and impair the functionality of the system base.
Observe the following:
•
Configuration with one column compartment
: Before moving the
system stack, unlock the system base using the locking tools.
•
Configuration with two or three column compartments or a Charger:
Before moving the system stack, dismount the system stack (for
instructions, see below).
Dismounting the system stack
Preparations
1. Turn off all modules with their main power switch.
2. Remove the solvent reservoirs and solvent lines from the solvent
rack.
3. Remove all flow connections between the modules within the
Vanquish system and all waste lines.
4. Remove all signal cables and power cords.
5. Grasp the system by its sides and slightly lift the column
compartment up and out of the grooves on the mounting bars.
6. Lift the front part of the solvent rack and remove the solvent rack by
pulling it towards the front.
7. Remove the detector, autosampler, and pump as described in the
next section.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...