Maintenance and Service
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Page 107
7.5
Lamps
This section describes the replacement and installation of the
•
Replacing the UV Lamp (Deuterium Lamp)
•
Replacing the VIS Lamp (Tungsten Lamp)
NOTICE—Damage to the device
Usage of lamps not supplied or approved by Thermo Fisher Scientific for
this detector may result in inferior performance, detector malfunction or
may damage the detector. Detector guaranty may be voided.
Use only lamps supplied or approved by Thermo Fisher Scientific for this
detector.
7.5.1
Replacing the UV Lamp (Deuterium Lamp)
When
•
The lamp is defective
•
After a defined amount of operating hours of the lamp as a
preventive measure
•
The lamp intensity is too low for demanding applications
•
The lamp intensity fluctuates causing artifacts in the baseline
Parts required
•
UV lamp
•
Optionally:
¨
Isopropanol
¨
Lint-free tissue
Preparations
1. Turn off the UV lamp.
2. Turn off the device.
Содержание Thermo Scientific Vanquish VC-D40
Страница 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 136: ...7 Maintenance and Service Page 136 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...
Страница 174: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...