7.  Connect the jacket cable (Figure 1) to the 

corresponding connector of the controller.

It is highly recommended that users routinely 
inspect glass vessels for damage, and 
replace the vessels if damage is observed.

Disassembling the Glass Bioreactor vessel

1.  While supporting the vessel with one hand, 

pull up on the agitator motor assembly (Figure 
1) to remove. Set the motor assembly aside, 
as it 

MUST NOT 

be washed.

2.  Uncouple the jacket cable (Figure 1). Loosen 

the hook and loop fastener (Velcro) straps on 
the heating jacket (Figure 1) and remove. Set 
the heating jacket aside, as it 

MUST NOT

 be 

washed.

3.  Remove any septa that are in place, and 

replace the septa fitting. Remove all filters and 
tubing from the headplate assembly (Figure 2).

4.  Unscrew the four headplate retaining screws 

(Figure 2).

5.  Lift the headplate assembly from the glass 

vessel. Some pressure may need to be 
applied to release the headplate seal. We 
recommend that a second operator support 
the vessel stand if this is required. Ensure 
that none of the installations of the headplate 
assembly (cooling coil, impeller shaft, etc.) hit 
the glass rim. 

CAUTION: DO NOT USE ANY 

SHARP TOOLS

 to separate the headplate 

assembly from the glass vessel.

6.  Remove the glass vessel from the stand. 

Store the glass vessel inverted (Figure 3) or on 
its side, if adequately supported, to prevent 
rolling. 

CAUTION:

 

DO NOT

 store the vessel 

with the base on the bottom. Because the 
base is not flat, the glass vessel can  

 

easily tip over and become damaged.

7.  Under the flange of the glass vessel is an 

O-ring for centering the headplate’s position 
with respect to the glass vessel (Figure 3). 
Ensure that the O-ring on the flange around 
the glass vessel (as received) is intact when 
removing the headplate for cleaning purposes.

8.  Depending on the cleaning procedure, the 

headplate assembly may be placed in the 
stand for washing (Figure 4), but ensure that 
all surfaces can be easily cleaned and that the 
used cleaning agents drain off.

Figure 3. Glass vessel with O-ring, 
stored inverted.

O-ring