-8-

Electrical Requirements

The unit construction provides protection against the risk of electric

shock by grounding appropriate metal parts . The protection may not

function unless the power cord is connected to a properly grounded

outlet. It is the user's responsibility to assure a proper ground connec-

tion is provided.

Refer to the serial number label on the rear of the unit to identify the specific
electrical requirements of your unit.

Make sure the voltage of the power source meets the specified voltage, ±10%.

Voltage Selection

(60 Hertz Units)

If the unit is to be operated from a 220 to 240 volt source, a voltage selector
inside the unit must be reset. See Section V, Service and Maintenance, for
instructions on changing the voltage selection.

NOTE Units rated 380-415 volt do not have a voltage selection switch.

Plumbing

Requirements

The plumbing connections are located on the rear of the unit and are labelled
FACILITY WATER and RECIRCULATING WATER. FACILITY WATER connec-
tions are 1 inch nominal FPT. RECIRCULATING WATER connections are 1½
inch nominal FPT. See illustration on page 20.

Before installing the unit to an instrument that previously used tap water as a
cooling fluid, flush the instrument several times to remove any rust or scale
that has built up. The manufacturer of the instrument should be able to
recommend a cleaning fluid for their equipment.

Connect the FACILITY WATER SUPPLY to the cooling water source.
Connect the FACILITY WATER RETURN to the drain.

Connect the RECIRCULATING WATER SUPPLY to the inlet of the instru-
ment being cooled. Connect the RECIRCULATING WATER RETURN to the
outlet of the instrument being cooled.

Flexible tubing, if used, should be of heavy wall or reinforced construction.
All tubing should be rate to withstand 80 psi at +40°C. Make sure all tubing
connections are securely clamped. Avoid running tubing near radiators, hot
water pipes, etc. If substantial lengths of tubing are necessary, insulation
may be required to prevent loss of cooling capacity.

Содержание System IV

Страница 1: ...ientific com tc Product Service Information Applications Notes MSDS Forms e mail Voice Info 800 258 0830 Thermo Scientific System IV Heat Exchanger Thermo Scientific Manual P N U00682 Rev 03 04 2013 I...

Страница 2: ......

Страница 3: ...ientific com tc Product Service Information Applications Notes MSDS Forms e mail Voice Info 800 258 0830 Thermo Scientific System IV Heat Exchanger Thermo Scientific Manual P N U00682 Rev 03 04 2013 I...

Страница 4: ...uhe Germany Tel 49 0 721 4094 444 Fax 49 0 721 4094 300 info tc de thermofisher com Building 6 No 27 Xin Jinqiao Rd Shanghai 201206 Tel 86 21 68654588 Fax 86 21 64457830 info china thermofisher com St...

Страница 5: ...ure 4 SECTION I Safety Warnings 5 SECTION II GeneralInformation Description 6 Specifications 6 SECTION III Installation Site 7 Facility Water Requirements 7 ElectricalRequirements 8 Voltage Selection...

Страница 6: ...ion 15 Voltage Selection 15 Fuses 15 PhaseRotation 15 PumpMotor Overload Protector 16 DisplayingSoftwareVersion 16 SECTION VI Troubleshooting ErrorCodes 17 Service Assistance 18 TechnicalSupport 18 SE...

Страница 7: ...hat shipped with the unit WEEE RoHS This product is required to comply with the European Union s Waste Electrical Electronic Equipment WEEE Directive 2002 96 EC It is marked with the following symbol...

Страница 8: ...have a warranty against defective parts and workmanship for one full year from date of shipment See back page for more details Out of Box Failure An Out of Box Failure is defined as any product that f...

Страница 9: ...leaking equipment Never operate the unit without cooling fluid in the reservoir Always turn off the unit and disconnect the line cord from the power source before performing any service or maintenance...

Страница 10: ...oads it may be because the system pressure is too low or the pump flow is too high Contact our Customer Support 4 Additional dimensions are shown on page 20 Section II General Information Description...

Страница 11: ...nit to operate at its full rated capacity If the cooling water does meet these standards the cooling capacity will be derated The chart is based on a 10 C difference between the temperature of the coo...

Страница 12: ...t and are labelled FACILITYWATER and RECIRCULATING WATER FACILITY WATER connec tions are 1 inch nominal FPT RECIRCULATING WATER connections are 1 inch nominal FPT See illustration on page 20 Before in...

Страница 13: ...f automotive anti freeze will void the manufacturer s warranty We recommend using distilled deionized water with a 0 05 to 0 1 MOhmcm reading Highly distilled deionized water above the 3 MOhmcm region...

Страница 14: ...ng on the solenoid valve to the make up fluid source using5 16 or 3 8 inch ID flexible tubing Tubing is available from Thermo Fisher Contact our Sales Department for more information see Preface After...

Страница 15: ...ted tripping of the monitor If the unit shuts down top off the reservoir and restart it When the system is full the reservoir level will no longer drop when the RECIRCULATING FLOW CONTROL valve is ope...

Страница 16: ...ng temperature remember that the lowest achiev able temperature is a function of the available flow rate the temperature of the cooling water supply and the heat load The temperature control system ac...

Страница 17: ...ss YES to displayCooL Press NEXT ENTER to display Stor press YES to accept all the changes pressNO to abort all changes and restore all the previous values The controller will again display the fluid...

Страница 18: ...nnect the power cord from the power source and turn off the facility cooling water Place a container under the strainer to collect any water that spills out of the basket when it is removed Unscrew th...

Страница 19: ...le 200 208V and 220 240V Set the switch for the appropriate range Replace the top cover Fuses The main fuses are 20 Amp RK5 The controller fuses are 1 Amp time delay Do not replace with an alternate a...

Страница 20: ...s display In the event the unit is not operational the software version and the check sum can also be read from the label which is on the microprocessor chip itself The following chart uses an example...

Страница 21: ...ff in the Setup Loop If Er16 is displayed the unit needs calibration contact us for assistance Display Indication Er 04 13 Interrupt error Er 14 Synchronous error Er 15 Asynchronous error Er 16 Bad ca...

Страница 22: ...er sale Support Before calling please obtain the following information Part number Serialnumber Voltage of unit Voltage of power source SoftwareVersion Technical Support Our Service Department can pro...

Страница 23: ...19 Section VII Diagrams Flow Diagram...

Страница 24: ...20 Rear View...

Страница 25: ...nized water Do not use untreated tap water as the total ionized solids level may be too high The desired level for long time usage is 1 to 3 MOhmcm compensated to 25 C The above recommendations will r...

Страница 26: ...r installation or maintenance or to any unit that has been operated or maintained in any way contrary to the operating or maintenance instructions specified in this Instruction and Operation Manual Th...

Страница 27: ......

Страница 28: ...Thermo Fisher Scientific 81 Wyman Street P O Box 9046 Waltham Massachusetts 02454 9046 United States www thermofisher com...

Отзывы: