Installation
Integrating the Detector into the LC System
14
Surveyor RI Plus Detector Hardware Manual
Thermo Scientific
Integrating the Detector into the LC System
Place the detector on a bench top as close as possible to the chromatographic column outlet
(minimizing the length of tubing necessary for connection to the flowcell inlet). If you are
using other Surveyor Plus modules, place the Surveyor RI Plus Detector on top of the
Surveyor Plus Autosampler as shown in
Be sure to place the HPLC system in a
draft-free
location away from an open window, air
conditioner vents, or other circulating air source. A stable room temperature (
±
0.5 °C) is
necessary for applications requiring maximum detection sensitivity. Allow at least 15 cm
(6 in.) of clear space between the back panel of the detector and any wall or obstruction. This
provides access to the back-panel connectors and a free flow of cooling air.
Figure 9.
Surveyor Plus LC system
Temp
Run
Comm
Power
Degas
Run
Comm
Power
Temp
Run
Comm
Power
Power switches
Detector
LEDs
Autosampler
LEDs
Pump
LEDs